The principle of operation
The principle of operation of the spectrometer is based on the analysis of the characteristic X-ray fluorescence spectra of radiation occurring in the measured sample under the influence of X-ray radiation.
The high-voltage power supply generator of the X-ray tube provides a stable supply voltage. X-ray tube radiation (primary X-ray radiation) irradiates the samples. The samples then emit fluorescent characteristic X-ray radiation. The fluorescent characteristic X-ray radiation of the sample is received by a semiconductor detector, amplified and fed into a multichannel pulse spectrometer compatible with a personal computer. Control of the spectrometer, selection of measurement modes and parameter control, processing of X-ray radiation spectra, obtaining measurement results in units of quantities characterizing the content of the component, or in intensities (relative intensities) of spectral lines of X-ray radiation is carried out using a computer using special programs.
General characteristics
Type: Energy Dispersive X-ray fluorescence Spectrometer
Model: ATM X RAY-X
Size: 650×565×390 mm;
Sample placement chamber size: Ø340×64(H) mm;
Weight: approx. 85 kg;
• Range of measured elements: from fluorine (F) to uranium (U) (depending on modification)
• Range of determination of the content of elements: ppm-99.99% (for different elements – different ranges)
• Simultaneous analysis of elements
• Measurement time: 60s - 200s
• Energy resolution of the detector: ≤129 eV
• Sample material: powder, solid, liquid
The measurement results in units of quantities characterizing the content of the component, or in the intensities (relative intensities) of spectral lines of X-ray radiation are carried out using a computer using special programs.
Configuration*
· High-performance X-ray tube with ultra-thin window
· Silicon Drift Detector SDD
· Digital controller ADC
· Advanced optical path
· High signal-to-noise ratio preamplifier
· Built-in high definition video camera
· Automatic switching of collimators and filters
· Carousel for multiple samples (with additional sample rotation system)
· Automatic correction of the spectrum drift
· Triple protection mode
· Vacuum Pump with power Cable
· USB cable, 2 pcs
· USB 2.0 to RS485 Universal Cable
· Suction bag
· Sample spoon
· Nylon tube
· Film in a roll (0.08*220 m, 4 microns
· Plastic pipette, (3 ml), 100 pcs
· Reusable plastic cups for solid samples, 12 pcs (in box 17)
· Disposable cups for liquid samples, 60 pcs (in box 63)
· Holders for cups for liquid samples, 12 pcs
· Steel rings for cups for liquid samples, 12 pcs
· Instructions on the USB disk
*depending on the modification
Additional features
· Single sample chamber and multi-sample chamber;
9/12/20-position carousel.
· Switching between vacuum and helium by pressing a button;
To measure the contents of light elements in powders or liquids, you can use a switch between vacuuming (for powders) and helium supply (for liquids).
· Universal intelligent testing system;
There is no need to choose different types of materials. Automatic detection of the material eliminates the human factor. This improves the accuracy of the analysis.
· Detector protection and electromagnetic blocking;
The protective plate protects the detector window from damage due to improper operation. An electromagnetic lock prevents the lid from opening during analysis and radiation leakage.
· Ultrashort optical path, ultra-low detection limit;
The ultrashort optical path and recommended filter combinations make it possible to determine the content of elements starting from F (fluorine), with a detection limit in light matrices for a number of elements of only 0.2 ppm.
· Accurate quantitative and semi - quantitative analysis;
ROHS5.0, Composite analysis V4.0, Coating analysis, Intelligent analysis of an unknown sample using the FP method.